Abstract
We investigate the use of Abbe-type refractometers with homogeneous absorbing fluids. We derive a simple model based on electromagnetic optics to describe quantitatively the angular distribution of light around the critical angle and report some experimental tests. First, we verify experimentally that the diffusing film near the base of the prism does in fact deliver light that refracts up to the critical angle. Then we compare experimental profiles of refracted light around the critical angle with theoretical profiles, and find very good agreement. Finally, we use the model to analyze some requirements and limitations on using Abbe-type refractometers on absorbing fluids.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.