Abstract
Osseointegration is a prerequisite for achieving a stable long-term fixation and load-bearing capacity of bone anchored implants. Removal torque measurements are often used experimentally to evaluate the fixation of osseointegrated screw-shaped implants. However, a detailed understanding of the way different factors influence the result of removal torque measurements is lacking. The present study aims to identify the main factors contributing to anchorage. Individual factors important for implant fixation were identified using a model system with an experimental design in which cylindrical or screw-shaped samples were embedded in thermosetting polymers, in order to eliminate biological variation. Within the limits of the present study, it is concluded that surface topography and the mechanical properties of the medium surrounding the implant affect the maximum removal torque. In addition to displaying effects individually, these factors demonstrate interplay between them. The rotational speed was found not to influence the removal torque measurements within the investigated range.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Journal of the Mechanical Behavior of Biomedical Materials
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.