Abstract

Several real-time RT-PCR assays have received Emergency Use Authorization from the United States Food and Drug Administration. The BD MAX™ SARS-CoV-2 assay, run by the BD MAX™ system, is a qualitative test that detects the SARS-CoV-2 specific nucleocapsid phosphoprotein gene regions, N1 and N2. The human RNase P gene is used as the endogenous nucleic acid extraction control. The Cepheid Xpert® Xpress SARS-CoV-2 assay, run by the GeneXpert system, detects the pan-sarbecovirus E gene and the N2 region of the N gene. We evaluated the performance characteristics of the BD and Cepheid assays using matched patient samples. We also analyzed comparative Ct values for both assays using 183 positive samples tested at this facility. In addition, we mitigated reporting false positive results without relying on interpretive software. We found that both systems showed comparable sensitivity. We found an approximately 3.5% false positive rate from the BD MAX™ system results.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.