Abstract
In the present study a QbD acquiescent method utilizing design of experiment and method operable design regions methodology was evaluated for optimization of chomatographic condition for separation of ten degradation products along with peaks of aspirin and lansoprazole. Box-Behnken design was exploited to evaluate the main and interaction effects of the selected critical process parameters on the critical quality attributes, viz resolution between thee peak pairs and retention time of last eluting peak. The optimal separation was predicted at pH 3.6, with a gradient starting at 20% of acetonitrile and initial hold time of 10 min. The results of experimental methods show excellent agreement with predicted results, highlighting the importance of design of experiments in method optimization. The selected working condition was then fully validated according to ICH guidelines for linearity, range, accuracy, precision and robustness.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.