Abstract

For the 7-term TRL calibration of a four-sampler vector network analyzer (VNA), expressions for the deviations of the S-parameters of two-port test objects are presented as functions of the deviations from the ideal values of the S-parameters of the TRL standards used. The sensitivity coefficients obtained are suitable for establishing the Type-B uncertainty budget for the S-parameter measurement.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.