Abstract
In the paper, dynamic calibration of air gauges with small chambers is discussed. Proposed sine input test rig is described, and its uncertainty is evaluated. Three main sources of uncertainty are identified as a Mechanical Unit, Pneumatic Unit and Digital Unit. Type B uncertainty estimation is considered, and the most difficult to assess sources are pointed out. Namely, it is the fluctuation of the back-pressure due to the airflow inside the air gauge. Thus, it was proposed to estimate the Type A uncertainty through the repetitions of the time constant measurement. The results provided highly satisfactory level of expanded uncertainty ca. 0.001 s. This result is covering all factors of influence on the dynamic calibration result.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.