Abstract

Quantum noise is a random process in X-ray-based imaging systems. We addressed and measured the uncertainty of radiomics features against this quantum noise in computed tomography (CT) images. A clinical multi-detector CT scanner, two homogeneous phantom sets, and four heterogeneous samples were used. A solid tumor tissue removed from a male BALB/c mouse was included. We the placed phantom sets on the CT scanning table and repeated 20 acquisitions with identical imaging settings. Regions of interest were delineated for feature extraction. Statistical quantities-average, standard deviation, and percentage uncertainty-were calculated from these 20 repeated scans. Percentage uncertainty was used to measure and quantify feature stability against quantum noise. Twelve radiomics features were measured. Random noise was added to study the robustness of machine learning classifiers against feature uncertainty. We found the ranges of percentage uncertainties from homogeneous soft tissue phantoms, homogeneous bone phantoms, and solid tumor tissue to be 0.01-2138%, 0.02-15%, and 0.18-16%, respectively. Overall, it was found that the CT features ShortRunHighGrayLevelEmpha (SRHGE) (0.01-0.18%), ShortRunLowGrayLevelEmpha (SRLGE) (0.01-0.41%), LowGrayLevelRunEmpha (LGRE) (0.01-0.39%), and LongRunLowGrayLevelEmpha (LRLGE) (0.02-0.66%) were the most stable features against the inherent quantum noise. The most unstable features were cluster shade (1-2138%) and max probability (1-16%). The impact of random noise to the prediction accuracy by different machine learning classifiers was found to be between 0 and 12%. Twelve features were used for uncertainty measurements. The upper and lower bounds of percentage uncertainties were determined. The quantum noise effect on machine learning classifiers is model dependent. • Quantum noise is a random process and is intrinsic to X-ray-based imaging systems. This inherent quantum noise creates unpredictable fluctuations in the gray-level intensities of image pixels. Extra cautions and further validations are strongly recommended when unstable radiomics features are selected by a predictive model for disease classification or treatment outcome prognosis. • We addressed and used the statistical quantity of percentage uncertainty to measure the uncertainty of radiomics features against the inherent quantum noise in computed tomography (CT) images. • A clinical multi-detector CT scanner, two homogeneous phantom sets, and four heterogeneous samples were used in the stability measurement. A solid tumor tissue removed from a male BALB/c mouse was included in the heterogeneous sample.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.