Abstract

A new primary frequency standard, the atomic caesium fountain CSF1, has been put into operation at the Physikalisch-Technische Bundesanstalt (PTB). 133Cs atoms are cooled in a magneto-optical trap and in optical molasses. They are launched to a height of 39 cm above the microwave cavity centre. The resulting Ramsey resonance signal has a full-width half-maximum linewidth (FWHM) of 0.88 Hz. The first uncertainty evaluation yields a relative 1 σ frequency uncertainty of 1.4 × 10−15. The short-term relative frequency instability of the CSF1 for averaging time τ is typically 3.5 × 10−13 (τ/s)−1/2, dictated by the available quartz oscillator as the local frequency source.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.