We present an uncertainty evaluation of a new cryogenic current comparator (CCC) bridge that is designed to compare two-terminal 1 and 10 MOmega standard resistors with the quantized Hall resistance (QHR) and scale from these decade values to other decade values between 10 kOmega and 1 GOmega. Use of the bridge over a wide range of resistance values results in a spectrum of dominant contributions to the uncertainty budget, and this analysis helps determine the optimum parameters of the measurements. Our theoretical analysis is compared with experimental data.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: IEEE Transactions on Instrumentation and Measurement
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.