Abstract

The aim of this work was to identify the main sources of uncertainty, quantify the standard deviation of each source of uncertainty, and calculated combined and expended uncertainties for the determination of linezolid in injectable dosage forms by UV spectrophotometry. The contributions of precision, linearity and weight of linezolid reference standard are the most significant, contributing with about 77% of the overall uncertainty. The uncertainties on the absorbances (sample and standard), volumetric flasks and volumetric pipettes have virtually no influence on the overall uncertainty. The estimated uncertainty of the spectrophotometric method for determination of linezolid was adequate for the scope of the method.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call