Abstract

A liquid flow standard system is used to calibrate liquid volume of fuel–oil flow meters at small flow rates between 50 L/h and 700 L/h. However, the system has not been used to calibrate volume flow rate because the system is only operated with the standing-start-and-finish mode. In this study, the liquid flow standard system was rebuilt to provide a calibration service of volume flow rate by attaching two flow diverters, which can operate the system with the flying-start-and-finish mode. To evaluate its performance for volume flow metering, several techniques were introduced. First, diverter timing errors were estimated by linear regression. Second, covariance between buoyancy correction factor and water density was obtained to consider interdependency between the two measurands. Third, calibration and measurement capability (CMC) was evaluated by setting a fixed value of collected weight or elapsed time for flow diversion. Finally, several CMCs were compared to find the best measurement condition. As a result of the above approach, the CMC of the liquid flow standard system was found to be (0.10–0.52)% (k = 2) for (50–700) L/h with a minimum collected weight at 10kg.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.