Abstract
The uncertainties in 60Co gamma-ray dosimetry at JAERI have been estimated by reviewing irradiation and calibration procedures. The uncertainty in dose delivered by the JAERI calibration system has been estimated to be 2.2% (2σ) and the uncertainty in dose measurement using JAERI alanine/ESR transfer dosimeters has been estimated to be 3.4% (2σ). A two-way dose intercomparison study between NPL and JAERI employing alanine-PS, alanine-paraffin, and dichromate dosimeters as transfer dosimeters was performed to confirm the estimated uncertainty over the dose range 1–50 kGy. Dose intercomparison results using both alanine and dichromate dosimeters show agreement within about 2%, which gives confidence in the estimated uncertainty in 60Co gamma-ray dosimetry at JAERI.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.