The Romanian standard for the analytical method for phenols measurements in aquatic media was subject of internal validation procedure, adapted to the specific concentration domain of the laboratory. Method performance criteria were investigated on standard samples measurements: selectivity, linearity domain, precision, accuracy, limit of detection, limit of quantification and robustness. The validated method was further used on real samples of domestic and industrial wastewaters from the Wastewater Treatment Plant of Brasov, Romania. As a novelty, even not required in the Romanian standard, sources of uncertainties were identified, and uncertainty was estimated for both measurements on standard and wastewater samples. The results showed that the method is suitable for phenolic compounds determination from wastewaters, with a very low uncertainty level.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.