Abstract

Ultrasound light diffraction in conjunction with a computer-tomographic evaluation procedure allows ultrasonic fields to be mapped as long as the limits of weak acousto-optic interaction are not exceeded. Beyond these limits, the light-sound interaction is far more complex, and the simplified evaluation based on the Raman-Nath approach will give rise to considerable errors. Two typical idealized sound beam configurations, of a circular and a square sound column, are used for numerically treating the problem. It could be shown that, depending on the light-sound interaction length and the ultrasound frequency, errors of the order of 50% are not rare.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.