Abstract

This paper presents a quantitative assessment of uncertainty for the 3D reconstruction of stents. This study investigates a CP stent (Numed, USA) used in congenital heart disease applications with a focus on the variance in measurements of stent geometry. The stent was mounted on a model of patient implantation site geometry, reconstructed from magnetic resonance images, and imaged using micro-computed tomography (CT), conventional CT, biplane fluoroscopy and optical stereo-photogrammetry. Image data were post-processed to retrieve the 3D stent geometry. Stent strut length, separation angle and cell asymmetry were derived and repeatability was assessed for each technique along with variation in relation to μCT data, assumed to represent the gold standard. The results demonstrate the performance of biplanar reconstruction methods is comparable with volumetric CT scans in evaluating 3D stent geometry. Uncertainty on the evaluation of strut length, separation angle and cell asymmetry using biplanar fluoroscopy is of the order ±0.2mm, 3° and 0.03, respectively. These results support the use of biplanar fluoroscopy for in vivo measurement of 3D stent geometry and provide quantitative assessment of uncertainty in the measurement of geometric parameters.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.