Abstract

We present two cases of unanticipated difficult airway in patients requiring reoperation after cervical spine instrumentation. In both cases, the upper airway examination was normal, and fiberoptic-guided intubation proceeded with the patient sedated and breathing spontaneously. Cord visualization was difficult, but the scope was eventually advanced into the trachea and the endotracheal tube placed safely. Later review of radiographs showed the previously unrecognized protrusion of cervical hardware into the meso- and hypopharynx. We recommend that anesthesiologists review recent radiographic studies for potential airway compromise before approaching the airway of patients presenting for revision of cervical instrumentation.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.