Abstract

This report is concerned with the investigation of double-focusing sector field inductively coupled plasma mass spectrometry (DF-ICPMS) for ultratrace and isotopic ratio analysis of long-lived radionuclides (226Ra, 230Th, 232Th, 233U, 237Np, 238U, and 241Am) using the direct injection high efficiency nebulizer (DIHEN). A new shielded torch arrangement, known as the guard electrode, improves relative sensitivity by a factor of six when the DIHEN is used. Absolute sensitivity with the DIHEN is on the order of 1300 (226Ra) to 1700 (238U) counts/fg at a solution consumption rate of 5 μL/min. This is a factor of from three to 20 better than the results obtained by a conventional nebulizer-spray chamber arrangement (e.g., ultrasonic and pneumatic nebulizers). The DIHEN-DF-ICPMS is successfully tested for isotope ratio measurements of 235U/238U standards and environmental radioactive waste solutions.