Abstract

Abstract We report the development of a fieldable frequency standard based on 199Hg+ ions confined in a hybrid r.f./dc linear ion trap. This trap permits storage of large numbers of ions with reduced susceptibility to the second-order Doppler effect caused by the r.f. confining fields. A 160 mHz wide atomic resonance line for the 40·5 GHz clock transition is used to steer the output of a 5 MHz crystal oscillator to obtain a stability of 2 × 10−15 for 24 000 s averaging times. For longer averaging intervals, measurements are limited by instabilities in available hydrogen maser frequency standards. Measurements with 37 mHz linewidth for the Hg+ clock transition demonstrate that the inherent stability for this frequency standard is at least as good as 1 × 10−15.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.