Abstract

BackgroundExternal ultrasound transducer disinfection is common practice in medicine. Unfortunately, clinically significant organisms, such as methicillin-resistant Staphylococcus aureus (MRSA), Pseudomonas aeruginosa, and Klebsiella pneumonia spread throughout healthcare facilities via direct contact despite disinfection protocols. Ultrasound transducers and coupling gel provide potential vectors for pathogen transmission, especially in immunocompromised and high-risk patient populations. Our objective was to conduct a survey to investigate the variety of cleaning solutions or sanitary wipes used and evaluate current standard practice for transducer disinfection across emergency medicine training programs in the United States.FindingsEighty-three academic emergency medicine programs participated in this study. Eighty-seven percent (95 % CI 80–94 %) of responding programs do not have a mandated protocol or standard contact time for transducer disinfection. Ninety percent (95 % CI 84–96 %) of institutions use disinfectant solution or disinfectant wipes, as the standard of practice, to cleanse ultrasound transducers after every use.ConclusionsCurrently, there is a great deal of variability with regard to non-endocavitary transducer disinfection protocols that seems to stem from the vast number of disinfectant products and ultrasound manufacturer disparate recommendations. In order to mitigate risk to patients and reduce health care costs linked to nosocomial infections; healthcare providers, ultrasound companies, and disinfectant manufacturers must develop a universal use disinfectant and a standard protocol for ultrasound device disinfection for noncritical device disinfection in the emergency department.Electronic supplementary materialThe online version of this article (doi:10.1186/s13756-016-0110-y) contains supplementary material, which is available to authorized users.

Highlights

  • External ultrasound (US) transducer disinfection is common practice in modern medicine

  • Currently, there is a great deal of variability with regard to non-endocavitary transducer disinfection protocols that seems to stem from the vast number of disinfectant products and ultrasound manufacturer disparate recommendations

  • In order to mitigate risk to patients and reduce health care costs linked to nosocomial infections; healthcare providers, ultrasound companies, and disinfectant manufacturers must develop a universal use disinfectant and a standard protocol for ultrasound device disinfection for noncritical device disinfection in the emergency department

Read more

Summary

Introduction

External ultrasound (US) transducer disinfection is common practice in modern medicine. No universal standard for external-use US transducer disinfection exists. Significant organisms, such as methicillinresistant S. aureus (MRSA), P. aeruginosa, K. pneumonia spread throughout healthcare facilities via direct contact despite disinfection protocols [1]. US transducers and coupling gel provide potential vectors for pathogen transmission, especially in immunocompromised and high-risk patient populations [2]. External ultrasound transducer disinfection is common practice in medicine. Clinically significant organisms, such as methicillin-resistant Staphylococcus aureus (MRSA), Pseudomonas aeruginosa, and Klebsiella pneumonia spread throughout healthcare facilities via direct contact despite disinfection protocols. Ultrasound transducers and coupling gel provide potential vectors for pathogen transmission, especially in immunocompromised and high-risk patient populations.

Objectives
Methods
Results
Discussion
Conclusion