Abstract

The purpose of this study is to assess the reliability of a new ultrasound imaging (USI) parameter and procedure for the evaluation of the size of the deep cervical flexor (DCF) muscles. Thirty-one healthy male subjects (21.6±2.5 years old) participated in this study. Two images of the diagonal dimension (DD) of the DCF of the subjects, at 1.5 fingerbreadths below the laryngeal prominence of the thyroid cartilage in a relaxed state, were taken on separate days with a 1-week interval. The intraclass correlation coefficient between the days was 0.82 (95% confidence interval, 0.65-0.91). The standard error of measurement and minimal detectable change were 0.8mm and 2.1mm, respectively. The results indicated that the DD may be used as a reliable USI parameter to measure the DCF thickness in healthy subjects.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.