Abstract

Microwave imaging promises high contrast between tumor and normal breast tissues, but its spatial resolution is limited. Here, we present a multimodality approach for high-resolution microwave imaging, where microwave image reconstruction is structurally guided by ultrasound imaging. The combined imaging concept is demonstrated using tissue phantom measurements obtained from a 16 x 15 transmitter/receiver microwave imaging system and a modified B-mode ultrasound system. With the geometry of the target and background known a priori from ultrasound, successful dielectric property images are recovered using a finite element-based reconstruction algorithm. We show that a target as small as 1.2 mm in diameter can be imaged with the multimodality approach, whereas it is impossible to detect such a small-size object using microwave imaging alone. The pilot clinical studies on two cases suggest that breast tumors can be much more accurately detected by the multimodality method.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.