Abstract
Fine-needle aspiration biopsy (FNAB) is a simple technique for the investigation of suspicious thyroid nodules. However, low success rates are reported in the literature. The aim of this prospective study was to compare the clinical performance and impact of an automatic aspirator, referred to here as Aspirator 3, to those of the manual technique for the FNAB of clinically suspicious thyroid nodules. One hundred nine consecutive patients with 121 clinically suspicious thyroid nodules underwent a biopsy twice of the same site with the clinically approved Aspirator 3 and with the manual technique. The number of follicular cell formations and the total number of follicular cells in the aspirate were counted using the ThinPrep® method. With the Aspirator 3, the total number and the mean number of extracted cell formations were significantly higher than the values achieved with the manual technique (total: 3222 vs. 1951, p=0.02; mean: 27 vs. 16). The total number of cells that were biopsied was also higher when the Aspirator 3 was utilized (47,480 vs. 23,080, p=0.005). Overall, the Aspirator 3 was superior in 65 biopsies, and the manual technique was superior in 39 biopsies. In terms of cell formations and the total number of cells aspirated, the Aspirator 3 was superior to the manual technique. Further, the Aspirator 3 was more convenient to use and had a greater precision in needle guidance.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.