Abstract
There is an accepted clinical requirement for a luting cement that can be command set upon satisfactory placement of an orthodontic appliance onto dentition. This work evaluates the suitability of ultrasound, imparted from a dental scaler, as a potential mechanism for achieving this. The net setting times and subsequent compressive strengths of a range of commercial and experimental glass polyalkenoate cements (GPCs) were evaluated, using modified ISO 9917 methods, when set both chemically and by ultrasound. The ultrasound was applied to the GPC through an orthodontic brace. It was possible to command set GPCs by the application of five to ten seconds of ultrasound; the exact time required being dependent upon the composition of the GPC in question. The compressive strengths of these cements can be improved by around 90% with the command set when the optimum PAA molecular weight and tartaric acid content is employed.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Journal of Materials Science: Materials in Medicine
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.