Abstract

PurposeTo study the propagation of ultrasonic shock waves in viscoelastic agents and the resulting corneal load. SettingUniversity Siegen, Institute for Mechanics and Control Engineering, Siegen, Germany. MethodsThe anterior chamber of a manufactured artificial eye was constructed according to anatomic dimensions. Three openings were drilled—for the phaco tip, for the exchange of a viscoelastic agent or water, and for the shock-wave sensor. The sensor was fixed to the area corresponding to the corneal apex. The sensor signal was analyzed using a direct oscilloscope that measured the amplitude reaching the corneal apex. Shock-wave propagation in several viscoelastic agents was compared with that in balanced salt solution. ResultsIn hydroxypropyl methylcellulose, the shock wave was amplified or influenced slightly. In hyaluronic-acid preparations, acoustic dampening occurred. ConclusionRemoval of hyaluronic-acid derivatives prior to phacoemulsification is not necessary.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.