Abstract
Dental implants can be rendered useless because of complications related to implant components such as the fracture of abutment screw and prosthetic retaining screw. Removal of such fractured fragments is therefore deemed necessary for the sake of prosthetic restoration. This case report presents a rare incident of healing abutment fracture and details the stepwise and probably the most conservative clinical technique using ultrasonic oscillations for retrieval of fractured screw fragment without affecting the internal surface of the implant. The management was based on the fact that the screw, being an inclined plane, will unthread when subjected to continuous vibrations.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.