Abstract

The purpose of this study is to measure blood flow velocity of rabbit retinal vessels using a 45-MHz ultrasonic Doppler system with a needle transducer. A high-frequency pulsed Doppler system that utilizes a 45-MHz PMN-PT needle transducer was developed to measure retinal blood flow velocity in situ. The pulsed Doppler allowed the differentiation of retinal from choroidal blood flow velocity. The needle transducer was inserted into the vitreous cavity through a 20-gauge incision port to access the retinal vessels. The first phase of the experiment evaluated the reproducibility of the measurements. The second phase measured velocities at four positions from the optic disc edge to the distal part of each vessel in nine eyes for the temporal and six eyes for the nasal portions. The angle between the transducer and the retinal vessel at each site was measured in enucleated rabbit eyes to estimate and compensate for measurement errors. In the first phase, the average measurement error was 5.97 +/- 1.34%. There was no significant difference comparing all eyes. In the second phase, the velocities gradually slowed from the disc edge to the distal part, and temporal velocities were faster than nasal velocities at all measurement sites. This study demonstrated the feasibility of reliably measuring retinal blood flow velocity using a 45-MHz ultrasonic Doppler system with a needle transducer.