Abstract

The fluid stream which emerges from an injection catheter of the type used in standard radiologic practice produces cavitation at high flow rates. Cavitation bubbles were detected within the aorta of dogs using an ultrasonic echo technique. Echo patterns produced by cavities in blood are characteristic. Fine bubbles produced by catheter injection occur only in the region of the effluent jet. Pressure recordings from the aorta proximal and distal to a rapid ejection show that pressure can drop below atmospheric in the region behind the jet flow. The effect of the injection on intravascular pressure is a function of the injection rate.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call