Abstract
Ultrasonic cleaning is often used as part of the manufacturing process of small medical devices such as guide wires and vascular implants. Ultrasonic cleaning at frequencies close to the natural frequency of the device can result in resonance, resulting in significant mechanical damage and possibly premature failure. This paper provides case studies of ultrasonic cleaning-induced fatigue and corresponding failures in small medical devices. Preventative measures, including analytical tools such as finite element analysis (FEA), to ensure that ultrasonic cleaning frequencies do not result in resonance and stresses sufficient to cause fatigue damage are also discussed.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.