Abstract

ABSTRACTUltrasonic-assisted solid-phase extraction via molecularly imprinted nanoparticles followed by HPLC-UV was used for trace determination of sotalol. Under optimized conditions, linear range, limit of detection, and limit of quantification of the proposed method were found to be 20–2000, 7.8, and 24 ng/mL, respectively. The adsorption capacity of sotalol for molecularly imprinted nanoparticles and non-imprinted nanoparticles was determined from the derived Langmuir isotherm and was found to be 31.55 and 15.22 mg/g, respectively. The proposed method was used for the separation of trace amounts of sotalol in human urine samples with acceptable high recoveries more than 94.40% and the inter-day precision with the coefficient of variation lower than 5%.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.