Abstract

As implant energies get lower and lower, significant errors can be present in junction depth measurements in secondary ion mass spectrometry (SIMS) ultrashallow depth profiling. Primary beam ion mixing is one of the main sources of errors leading to overestimation of junction depths in SIMS measurements. In this article, we systematically study the correlations between the implant profile trailing edge, junction depth and primary ion beam energy for low energy boron and arsenic implants. Using a mathematical deconvolution model proposed by Yang and Odom [Mater. Res. Soc. Symp. Proc. 669, J4–16 (2001)], we are able to estimate the error of the junction depth and consistently improve the accuracy of junction depth measurements using SIMS.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.