Abstract
Secondary ion mass spectrometry (SIMS) depth profiling is an important technique for the characterization of ultra shallow junctions, thin gate oxides and other interfacial layers in modern wafer fabrication. In this study, a CAMECA IMS 6f SIMS instrument is used to study the various factors that may affect the depth resolution in ultra shallow depth profiling of a B delta doped Si standard sample. Several analyses using 0.5-2.0 keV O<SUB>2</SUB><SUP>+</SUP> have been performed with and without oxygen flooding and sample rotation. The roughening of the sputtered crater bottom is one factor that degrades the depth resolution in low energy depth profiling, and both oxygen flooding and sample rotation are shown to suppress surface roughening. In addition, the depth resolution is found to be sensitive to other factors such as ion beam mixing and as a result, impact energies as low as 0.5 keV may be needed to achieve optimum depth resolution.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.