Abstract

An innovative method for the ultrasensitive detection of mercury by solution anode glow discharge atomic emission spectroscopy (SAGD-AES) coupled with hydride generation (HG) was first investigated. In this method, the mercury vapor generated by the HG was transmitted to the SAGD through the miniature hollow tungsten tube for excitation and detected by a miniaturized spectrograph. A thorough parametric evaluation of the HG and SAGD system was performed, including the type and concentration of carrier acid, He flow rate, concentrations of NaBH4, discharge current and discharge gap. Under optimal operating conditions, the detection limit for Hg2+ achieved 0.03 μg/L, with a relative standard deviation of 1.1% at the Hg2+ concentration of 5 μg/L. Moreover, the correlation coefficient of the calibration curve was 0.9996 in the range between 0.1 and 10 μg/L. The accuracy and practicability of HG-SAGD-AES were verified by measuring GBW09101b (human hair), GBW10029 (fish), soil and rice samples. The results showed good agreement with the certified values and values from direct mercury analyzer (DMA).

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.