Abstract
Picogram‐level gatifloxacin was determined based on its significantly catalyzed effect on myoglobin‐luminol chemiluminescence (CL) reaction in the flow injection system. The enhanced chemiluminescence intensity was linear with gatifloxacin concentration in the range from 50 ngl−1–10 µg l−1 (r2=0.9995), and the detection limit was 20 ng l−1 (3σ). At a flow rate of 2.0 ml min−1 for each line, a complete analytical process could be performed within 0.5 min, including sampling and washing, with a relative standard deviation of less than 4.0% (n=7). The proposed method was applied successfully in the determination of gatifloxacin in tablets, human serum and urine samples with the recovery from 97.4–104.5%.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.