Abstract
Hg2+ toxicity is one of the most common chemical poisonings that occurs mainly from drinking polluted water. In the current work, Phycocyanin (PC) was exploited as a fluorescent sensor for sensitive and selective detection of Hg2+ in an aqueous system. PC-Hg2+ interaction was monitored using a spectro-fluorometer under different buffered solutions at pH values of 6,7,8,9, or 10 above the isoelectric point of PC (5.18). A remarkable decrease of PC fluorescence intensity was observed under Tris-buffer at pH 6 upon the addition of increasing Hg2+ concentrations (1–120 nM). Under the maintained experimental conditions, the current sensor showed a good linear relationship with R2 = 0.9971 and a limit of detection as low as 0.7 nM was achieved. In addition, a notable selectivity of Hg2+ over other nine heavy metals (Cu2+, Zn2+, Pb2+, Mg2+, Mn4+, Li+, Fe3+, Co2+, and Al3+) was achieved in the presence of 120 nM of each metal. Moreover, the current fluorescent detection assay was also tested in real samples of pond water, and recoveries as well as relative standard deviations within the acceptable limits were recorded.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Journal of Environmental Science and Health, Part A
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.