Abstract

A 2 nV/ $\surd $ Hz and 10 fA/ $\surd $ Hz at 1-Hz ultralow noise preamplifier is designed to improve the sensitivity of the induction magnetometer (IM) from 0.1 Hz to 10 kHz. Meanwhile, the noise characteristics of IM are analyzed and distinguished. A novel optimization method to find the minimum length of the core under the required noise level and the limited IM weight is proposed. Furthermore, an optimal IM prototype from 0.1 Hz to 10 kHz is developed with the noise level of 1 pT/ $\surd $ Hz at 1 Hz and 1.5 fT/ $\surd $ Hz at 2 kHz. The length is only 360 mm and the weight is 1.5 kg. The measurement results of the IM also confirm the theoretical noise model and the optimization method.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.