Abstract
We demonstrate an apparatus for measuring time-dependent x-ray diffraction. X-ray pulses from a synchrotron are diffracted by a pair of Si(111) crystals and detected with an x-ray streak camera that has single-shot resolution of better than 1 ps. The streak camera is driven by a photoconductive switch, which is triggered by 100-fs laser pulses at a repetition rate of 1 kHz. The laser and the streak camera are synchronized with the synchrotron pulses. In the averaging mode, trigger jitter results in 2-ps temporal resolution. We measured the duration of 5-keV pulses from the Advanced Light Source synchrotron to be 70ps.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.