Abstract
In order to fully exploit the ballistic potential of particle therapy, we propose an online range monitoring concept based on time-of-flight (TOF)-resolved prompt gamma (PG) detection in a single proton counting regime. In a proof of principle experiment, different types of monolithic scintillating gamma detectors are read in time coincidence with a diamond-based beam hodoscope, in order to build TOF spectra of PG generated in a target presenting an air cavity of variable thickness. Since the measurement was carried out at low beam currents (< 1 proton/bunch) it was possible to reach excellent coincidence time resolutions, of the order of 100 ps (σ). Our goal is to detect possible deviations of the proton range with respect to treatment planning within a few intense irradiation spots at the beginning of the session and then carry on the treatment at standard beam currents. The measurements were limited to 10 mm proton range shift. A Monte Carlo simulation study reproducing the experiment has shown that a 3 mm shift can be detected at 2σ by a single detector of ∼1.4 × 10−3 absolute detection efficiency within a single irradiation spot (∼108 protons) and an optimised experimental set-up.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.