Abstract

We propose a novel technique for measuring directly in real time, locally and non-invasively the pulse wave velocity (PWV) on peripheral arteries. Very high frame rate ultrasonic imaging (> 1000 frames/s) was achieved for tracking in 2D the propagation of transient vibrations along arterial wall. The arterial pulse waves are observed within a single cardiac cycle allowing the estimation of the pulse wave velocity with a good accuracy. In this study, this technique was validated for PWV evaluation on 25 healthy patients using conventional sonographic probes. The mean carotid PWV was found to be 5.5 ± 1.2 m/s (from 4.5 m/s to 7 m/s) with good intra and interobserver variability (inferior of 10% of the mean). These data suggest a good accuracy and reproducibility of the technique for real time PWV evaluation in vitro and on healthy volunteers.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.