Abstract

Ultrafast breath-hold contrast material-enhanced magnetic resonance (MR) angiography can be performed with a flexible imaging sequence. With the current generation of high-speed imaging gradients, it is possible to achieve sequence repetition times of 4 msec or less. These repetition times make it possible to obtain high-resolution (512 x 512 x 64) images in under 30 seconds. Applications of this versatile technique include imaging of aortic dissection, thoracic and abdominal aortic aneurysm, pulmonary embolus, carotid stenosis, and peripheral vascular disease. The administration of contrast material must be tailored to the vascular anatomy under examination to avoid venous enhancement. The rapid data acquisition times can be used to image multiple temporal phases or multiple locations. With this technique and administration of a T1-shortening contrast agent, high-quality MR angiography can be routinely performed in a variety of vascular regions (eg, thoracic and abdominal aorta, pulmonary arteries, carotid arteries, lower extremities).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.