Abstract

To evaluate the feasibility of ultra-widefield retinal imaging in patients with near infrared (IR)-transmitting black intraocular lenses (IOLs). Oxford Eye Hospital, Oxford, United Kingdom. Laboratory evaluation of a diagnostic technology with interventional case report. The field of retinal imaging through a Morcher poly(methyl methacrylate) (PMMA) black IOL was determined in a purpose-built adult schematic model eye with the HRA2 Spectralis confocal scanning laser ophthalmoscope using standard imaging, Staurenghi retina lens-assisted imaging, and ultra-widefield noncontact imaging. Retinal imaging using each modality was then performed on a patient implanted with another Morcher PMMA black IOL model. Ultra-widefield noncontact imaging and lens-assisted imaging captured up to 150 degrees of field (versus 40 degrees with a standard confocal scanning laser ophthalmoscope). Ultra-widefield retinal images were successfully acquired in a patient eye with a black IOL. This study has identified the first ultra-widefield retinal imaging modalities for patients with near IR-transmitting black IOLs. Should larger studies confirm this finding, noncontact ultra-widefield confocal scanning laser ophthalmoscopy might be considered the gold standard imaging technique for retinal surveillance in patients with near IR-transmitting black IOLs. No author has a financial or proprietary interest in any material or method mentioned.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call