Abstract

Background levels of platinum were determined in human autopsy tissues taken from five individuals. The investigated specimens were lung, liver and kidney. Sample preparation involved microwave digestion followed by an open vessel treatment. Inductively-coupled plasma sector field mass spectrometry (ICP-SFMS) was applied in combination with an ultrasonic nebulization/membrane desolvation system for sample introduction. Isotope dilution analysis was employed for accurate quantification of platinum. Excellent procedural detection limits (3 s validation) of 20, 20 and 34 pg g(-1) dry weight were obtained for lung, liver and kidney tissue, respectively. Due to the lack of appropriate biological reference material, road dust (BCR-723) was used for method validation. Platinum levels ranging between 0.03 and 1.42 ng g(-1) were determined in the investigated samples. The platinum concentrations observed in human lung tissue may reflect the increasing atmospheric background levels of platinum originating from car catalysts. The presence of platinum in kidney and liver tissue samples clearly indicates the bioavailability of the element.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.