Abstract
A method has been developed for determination of twenty-four polar pesticides—nine organophosphorus pesticides, thirteen organonitrogen compounds, and two triazine degradation products—in surface water. It entails extraction of the target pesticides from 1-L water samples by solid-phase extraction (SPE), then gas chromatography (GC) with large-volume (40 μL) injection. Filtered surface water, from the St Lawrence River in Canada and the River Loire and its tributaries in France, was extracted on cartridges filled with 500 mg Carbopack B (120–400 mesh). Analysis was performed by gas chromatography with a thermionic specific detector (GC-TSD) and a mass spectrometric (MS) detector. Overall percentage recoveries were satisfactory (>70%) for all target pesticides, with precision below 10%. Detection limits were between 0.5 and 4 ng L−1.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.