Abstract

1.5T gradient echo-based R2∗ estimates are standard-of-care for assessing liver iron concentration (LIC). Despite growing popularity of 3T, echo time (TE) limitations prevent 3T liver iron quantitation in the upper half of the clinical range (LIC ⪆20 mg/g). In this work, a 3D radial pulse sequence was assessed to double the dynamic range of 3T LIC estimates. The minimum TE limits the dynamic range of pulse sequences to estimate R2∗. 23 chronically-transfused human volunteers were imaged with 1.5T Cartesian gradient echo (1.5T-GRE), 3T Cartesian gradient echo (3T-GRE), and 3T ultrashort TE radial (3T-UTE) pulse sequences; minimum TEs were 0.96, 0.76, and 0.19 ms, respectively. R2∗ was estimated with an exponential signal model, normalized to 1.5T equivalents, and converted to LIC. Bland-Altman analysis compared 3T-based estimates to 1.5T-GRE. LIC by 3T-GRE was unbiased versus 1.5T-GRE for LIC ≤ 25 mg/g (sd = 9.6%); 3T-GRE failed to quantify LIC > 25 mg/g. At high iron loads, 3T-UTE was unbiased (sd = 14.5%) compared to 1.5T-GRE. Further, 3T-UTE estimated LIC up to 50 mg/g, exceeding 1.5T-GRE limits. 3T-UTE imaging can reliably estimate high liver iron burdens. In conjunction with 3T-GRE, 3T-UTE allows clinical LIC estimation across a wide range of liver iron loads. Magn Reson Med 79:1579-1585, 2018. © 2017 International Society for Magnetic Resonance in Medicine.