Abstract

Elagolix (ELX) is an orally administered non-peptidic GnRH antagonist that has been approved by the Food and Drug Administration in 2018 for the treatment of endometriosis pain. A sensitive and selective method for estimating elagolix (ELX) in human plasma and content uniformity was developed and validated. The spectrofluorimetric technique was used to investigate ELX utilizing boron-doped carbon quantum dots (B@CQDs). After gradually adding ELX, the quantum dots fluorescence was enhanced with LOQ of 1.74 ng mL−1, the calibration curve between ELX and corresponding fluorescence intensity was found over a range of 4–100 ng mL−1. The method was successfully applied in real human plasma with pharmacokinetic study and content uniformity test. The pharmacokinetic parameters as Cmax were found to be 570 ± 5.32 ng. mL−1 after 1 h, t1/2 was found to be 6.50 h, and AUC was found to be 1290 ± 30.33 ng. h. mL−1. B@CQDs were characterized using variety of instruments. The strategy is simple to implement in clinical labs and therapeutic drug monitoring systems.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call