Abstract
This study aims to provide a rapid, sensitive and precise UPLC-MS/MS method for target steroid quantitation in biological matrices. We developed and validated an UPLC-MS/MS method to simultaneously determine 16 steroids in plasma and tissue samples. Ionization sources of Electrospray Ionization (ESI) and Atmospheric Pressure Chemical Ionization (APCI) were compared in this study by testing their spectrometry performances at the same chromatographic conditions, and the ESI source was found up to five times more sensitive than the APCI. Different sample preparation techniques were investigated for an optimal extraction of steroids from the biological matrices. The developed method exhibited excellent linearity for all analytes with regression coefficients higher than 0.99 in broad concentration ranges. The limit of detection (LOD) was from 0.003 to 0.1ng/mL. The method was validated according to FDA guidance and applied to determine steroids in sea lamprey plasma and tissues (fat and testes) by the developed method.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.