Abstract

In this work, we demonstrate the utility of ultra performance liquid chromatography-atmospheric pressure photoionization-tandem mass spectrometry (UPLC-APPI-MS/MS) for high-sensitivity and high-throughput analysis of United States Environmental Protection Agency (U.S. EPA) 16 priority pollutants polycyclic aromatic hydrocarbons (PAHs). Analyses were performed on a Waters Acquity-TQD equipped with Syagen's PhotoMate APPI source. All 16 PAHs were analyzed on column in approximately 3.5 min with excellent chromatographic separation for all PAH isomers and with low picogram detection limits on column for all analytes using chlorobenzene as a dopant. Dynamic linear ranges were evaluated and found to cover at least 3-4 orders of magnitude. In comparison with the existing U.S. EPA methods, this approach improves instrument sample throughput by at least 10-fold.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.