Abstract

A simple, sensitive and selective ultra-performance liquid chromatography method for the determination of voriconazole in plasma and oral fluid was developed and validated. After a liquid-liquid extraction with methyl-tert-butyl ether, the analyte and internal standard were separated on a Hypersil Gold C18 column (2.1 × 100 mm, p.d. 1.9 µm), eluted with a mobile phase composed of thietylammonium phosphate buffer and acetonitrile (70:30, v/v). Total run time was 4 min, total mobile phase consumption of 2.2 mL. Detection was performed with a photodiode array detector with quantitation at 256 nm. Voriconazole concentrations in oral fluid were on average 57.5% (±5.3) of those measured in paired plasma samples.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.