Abstract
For the first time, a dosimeter employing two floating gate radiation field effect transistors (FGRADFET) and operating at mere 0.1 V is presented. The novel dosimeter requires no power during irradiation and consumes only 1 during readout. Besides the low power operation, structural changes at the device level have enhanced the sensitivity of the dosimeter considerably as compared to previous designs. The dosimeter is integrated with a wireless transmitter chip, thus eliminating all unwanted communication and power cables. It has been realized monolithically in DALSA's 0.8 complementary metal-oxide-semiconductor process and characterized with X-ray and γ-ray sources. A maximum sensitivity of 5 mV/rad for X-rays and 1.1 mV/rad for -rays have been achieved in measurements. Due to its small size, low-power, and wireless operation, the design is highly suitable for miniaturized, wearable, and battery operated dosimeters intended for radiotherapy and space applications.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: IEEE Journal on Emerging and Selected Topics in Circuits and Systems
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.