Abstract

A simple and highly sensitive method for the determination of caffeine in energy beverages was developed and validated. Sample preparation utilizing solid-phase extraction (SPE) was simple and reliable. Separation by isocratic ultra-high performance liquid chromatography (UPLC) with a reversed-phase C18 column was performed within 6 min. The use of SPE with UPLC coupled with electrospray ionization-multiple tandem mass spectrometry detection (ESI-MSn) was accurate, reproducible, and validated for the determination of caffeine in energy drink matrices. The limit of quantification for caffeine was approximately 2.1 ng mL−1. The intra-day and inter-day precisions were less than 4%, and the accuracy of the measurements was between 85.1% and 93.2%. Results for caffeine concentrations in eighteen beverages were compared to the values on the labels. This paper describes the first use of the UPLC–ESI-ion trap MSn technique for quality-control purposes of caffeine present in energy drinks.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.