During the study, intraoral periapical images were analyzed in 300 patients. The possibilities of using the methods of radiography and their influence on the quality of the obtained x-ray images were studied. The intraoral periapical radiography was compared with the results of orthopantomography and cone beam computed tomography. To identify the features of the mapping of zones of periapical destruction, 47 experimental x-ray studies were performed on skeletonized jaws with artificial defects in cortical plates. Often encountered errors are: an arbitrary choice of angles of inclination and the centration of the x-ray tube, the wrong location of the x-ray receiver in the patient's mouth, inaccurate installation or poor fixation of the patient's head, inefficient selection of physical and technical conditions of shooting, non-compliance with the conditions of the photo process with analog radiography or post-processing and printing digital image. The characteristic projection distortions of images resulting from these errors are analyzed and illustrated.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.